Drew
Scientific 2280 |
||||||||||||||||||||||||
|
Specifications: Sensors: Laser diode illuminated flow cytometer with three, solid state sensors and one PMT tube for WBC differential WBC aperture: 100 µm diameter x 75µm length RBC/Pit aperture: 78µm diameter x 55µm length The Plt/RBC channel features the patented "von Behrens" Silencer which eliminates RBC recirculation that interferes with low Plt counts in conventional aperture designs. Dimensions (analyzer): 41.0 cm (16.5 in) Height x 58.0 cm (22.8 in) Width x 58.0 cm (22.9 in) Depth System weight: 325 Kg (72 lb) Display: Windows XP interface Power: The system works on 90 to 250 VAC 47 to 63 Hz, with no manual adjustment required. Operating conditions: The 2280 can work in wide range of working environments with a temperature range of 15°C to 32°C (60°F to 90°F) and relative humidity of 10% to 90% noncondensing. Ports: Two RS232 serial ports and one parallel printer port (Centronics) Printer: Windows XP supported, color or B/W printer can be used. |
|
The technology used combines to give a full CBC with 5 part differential in under one minute. Features
Sample Handling: Blood samples must be collected into EDTA tubes. Depending on how the 2280 is operated, the system requires: Direct mode: 180µL whole blood Sample saver mode: 80µL whole blood Automatic sampler**: 180 µL whole blood
Parameters: The system aslo displays: 2 x WBC scattergrams Plt histogram, RBC histogram. The instrument has a throughput of a minimum 80 samples per hour in direct sampling model.
Interested in buying
or learning more about this piece of equipment?
We also
carry these other
D3 Automatic Hematology System
|
||||||||||||||||||||||
![]()
|
|||||||||