YSI 2300 STAT Plus Glucose & Lactate Analyzer
The
YSI 2300 STAT PlusTM
is the only dedicated glucose and lactate analyzer available. This
analyzer makes precise glucose measurements in whole blood, plasma
or serum; precise lactate measurements in whole blood, plasma or
cerebrospinal fluid. The analyzer's hematocrit correction option
allows you to have glucose results expressed as plasma even when
running whole blood.
Speed
 |
Results in one
minute or less |
Low cost per test
 |
The immobilized enzymes do not have to be renewed after each
test, minimizing the use of costly reagents and supplies
|
Proven Technology
 |
You can be confident that results are accurate
|
Precise, accurate
results
 |
YSI provides the highest precision and most accurate results for
whole-blood glucose and lactate.
|
Whole-blood
capability
 |
The YSI 2300 is the only dual-channel glucose and lactate
analyzer that uses whole blood, so you do not waste spinning
blood. |
Small sample size
(25 µL)
 |
The analyzer requires a small sample, making it practical in
neonate applications |
Simple, safe
operation
 |
Menu-driven software guides operators through each step.
Diagnostics provide performance analysis and troubleshooting.
Samples are introduced to the analyzer by tubes or syringe; no
needles required. |
Batch operation
 |
The 24-position
turntable lets you run batches. You can assign each batch and
sample an identification number with date and time.
|
Autocalibration
 |
Calibration is automatic, so the analyzer is always ready.
|
Interference
rejection
 |
YSI immobilized enzyme technology eliminates most of the
interferences that plague other methods.
|
Broad measurement
range
 |
Up to 900 mg/dL glucose and 30 mmol/L lactate virtually
eliminates the need for sample dilution.
|

|
|
|
|
YSI 2300 STAT Plus
Glucose & Lactate Analyzer
Specifications:
Sample Size: 25 µL
(aspirated volume)
Respose Time: (from test tube holder)
Normal mode: Displayed and printed results in 65 seconds
Sample to sample interval is 100 seconds.
Screen Mode: Displayed and printed result in 45
seconds.
Measurement Range:
Normal mode: Glucose: to 900 mg/dL (9000 mg/L, 50.0
mmol/L)
Lactate: to 30.0 mmol/L (267 mg/dL, 2670 mg/L)
Screen Mode: Glucose: to 500 mg/dL
(5000 mg/L, 27.8 mmol/L)
Lactate: to 15.0 mmol/L (134 mg/dL, 1335 mg/L)
Precision:
Glucose: Whichever is larger, ±2% of reading or 2.5 mg/dL
(25 mg/L or 0.2 mmol/L)
Lactate: Whichever is larger, ±2% of
reading or 0.1 mmol/L
(1 mg/dL or 10 mg/L)
Calibration Point:
Glucose: Whichever is larger, ±2% of reading or 2.5
mg/dL
(25 mg/L or 0.2 mmol/L)
Lactate: Whichever is larger, ±2%
of reading or 0.1 mmol/L
(1 mg/dL or 10 mg/L)
Calibration Point:
Glucose: 180 mg/dL (1800 mg/L, 10.0 mmol/L)
Lactate: 5.00 mmol/L (45 mg/dL, 445
mg/L)
Membrane Working Life (typical)
Glucose: 21 days
Lactate: 14 days
Operating Temperature: 15 to 35°C
Humidity: 10 to 90% noncondensing
Power: 110-120 VAC or 220-240 VAC, 50-60 Hz
Size:
25.4 x 35.6 x 35.6 cm, 11.4 kg
10 x 14 x 14 inches, 25 lbs
Regulatory Compliance: CE, FDA
Registered Class II (862.1345), CSA
CE Device Category: IVD