Technical Specifications
Model sizes/Description | 110 | 260 | 410 | 750 | 1060 | 1400 | 2200 | ||
Stainless steel interior
|
Volume approx. l | 108 | 256 | 384 | 749 | 1060 | 1360 | 2140 | |
Width (A) mm | 560 | 640 | 1040 | 1250 | 1972 | ||||
Height (B) mm | 480 | 800 | 1200 | 1450 | |||||
Depth (less 32 mm for fan – Peltier) (C) mm | 400 | 500 | 600 | 850 | 750 | ||||
Max. number of grids/shelves number | 5 | 9 | 14 | 28 | 42 | ||||
Max. loading per grid/shelf kg | 20 | 30 | 20 | 30 | |||||
Max. loading of chamber kg | 150 | 200 | 250 | 330 | |||||
Max. loading per slide-in drip tray kg | 3 | 4 | 8 | - | |||||
Max. loading per bottom drip tray kg | 3 | 4 | 8 | - | |||||
Textured stainless
steel exterior |
Width (D) mm | 745 | 824 | 1224 | 1435 | 2157 | |||
Height (sizes 410, 750, 1060, 1400 and 2200 with castors) (E) mm | 864 | 1183 | 1720 | 1913 | |||||
Depth (without door handle, depth of handle +56 mm) (F) mm | 555 | 655 | 755 | 1005 | 905 | ||||
Standard equipment
|
Stainless steel grids, electropolished | 2 | 4 | 6 | |||||
Water tank including connection hose | 2.5 litres | 10 litres | |||||||
Standard works calibration certificate (measuring point chamber center) | 10, 37 and 30 °C/60 % rh | 25 °C/40 % rh and 40 °C/75 % rh | |||||||
Temperature
|
Working temperature range without light, without humidity °C | 0 (at least 20 below ambient temperature) to +70 | 15 (at least 10 below ambient temperature) to +60 |
||||||
Working temperature range without light, with humidity °C | 5 (at least 20 below ambient temperature) to +70 | 15 (at least 10 below ambient temperature) to +60 |
|||||||
Working temperature range with light, without or with humidity °C | 15 to 40 | - | |||||||
Setting temperature range without light, with humidity °C | 5 to 70 | 15 to 60 | |||||||
Setting temperature range with light, with humidity °C | 5 to 70 | - | |||||||
Setting temperature range with light, without humidity °C | 0 to 70 | - | |||||||
Setting temperature range without light, without humidity °C | 0 to 70 | 15 to 60 | |||||||
Setting accuracy °C | 0.1 | ||||||||
Humidity
|
Setting range humidity with light % rh | 10 to 85 | - | ||||||
Setting range humidity without light % rh | 10 to 90 | 10 to 80 | |||||||
Setting accuracy % rh | 0.5 | ||||||||
Further data
|
Electrical load at 230 V, 50/60 Hz approx. W | 420 | 700 | 1400 | 2000 | ||||
Electrical load at 115 V, 50/60 Hz approx. W | 420 | 700 | 1400 | - | |||||
Peltier elements in the rear number | 1 | 2 | 4 | 6 | |||||
Packing data
|
Net weight approx. kg | 77 | 122 | 160 | 208 | 260 | 450 | 493 | |
Gross weight (packed in carton) approx. kg | 102 | 173 | 213 | 279 | 424 | 639 | 730 | ||
Width approx. mm | 830 | 930 | 1330 | 1370 | 1560 | 2300 | |||
Height approx. mm | 1050 | 1380 | 1930 | 1910 | 1970 | 2200 | 220 | ||
Depth approx. mm | 800 | 930 | 1050 | 1300 | 1190 |