Technical Specifications
Mass Selective Detector
Mode (standard) | EI |
Modes (optional) | PCI, NCI, EI with CI source |
Ion source type | Noncoated inert EI source for turbomolecular pump systems
(optional for diffusion pump system) Stainless steel EI source for diffusion pump system |
Mass filter | Monolithic hyperbolic quadrupole |
Maximum mass | 1050 u |
Detector | Triple-Axis HED-EM with extended life EM |
Scan rate (electronic) | Up to 12,500 u/s |
Pumping system |
65 L/s diffusion pump, 70 L/s or 262 L/s turbomolecular
pump with 2.5m³/h mechanical pump
|
Gas Chromatograph (6850, 6890N, or 7890A GC)
For more specifications on GCs refer to the GC Data Sheet
Injector | Split/ splitless (standard), others available |
Autosampler | G2880A, 7683, CombiPAL, or 1888A (and more...) |
Oven temperature | Ambient +4 °C - 450 °C(6890/ 7890A) or +5 °C -350 °C (6850) |
Oven ramps/plateaus | 6850 and 6890: 6/7; 7890A: 20/21. Negative ramps are allowed. |
Electronicpneumatic
control(EPC) |
Auto pressure regulation for split/splitless, septum purge |
Carrier gas control modes | Constant pressure and flow modes, pressure and flow programmable |
Pneumatic splitter |
Capillary Flow Technology devices for effluent splitting backflushing,
and column switching
|
Data System
eMethods | Transfer methods between 5975 and 5973 Series MSDs |
Simultaneous MS and GC |
Six signals for the 7890A GC systems and four signals
for the 6890A Series GC systems (up to two MS) detector
data acquisitions
|
SIM/Scan |
Autmated SIM setup and high performance
synchronous SIM/scan operation
|
Application autotunes |
One-click autotune for BFB, DFTPP
|
Spectral libraries (optional) |
NIST, Wiley, Pfleger-Mauer Drug, Stan pesticide
|
Spectral DRS and RTl
databases (optional) |
Pesticides and endcrine disrupters, volatiles, PCBs,
toxicology, FAMEs, flavors, organotin compounds,
hazardous chemicals, indoor air toxics, Japan Positive
List, forensic toxicology and environment semivolatiles
|
21CFR11 Compliance |
Optional software available
|
Installation Checkout Specifications
EI scan sensitivity | 1-uL injection of a 1-pg/uL OFN standard scanning from
50 to 300 u will give a nominal m/z 272 ion With Helium: S/N 400:1 for turbo systems, 200.1 for the diffusion system With Hydrogen: S/N 100.1 for turbo systems and 50:1 for the diffusion system |
PCI scan sensitivity | 1-uL injection of a 100-pg/uL BZP standard scanning
from 80 to 230 u will give 125:1 S/N with Methane and 500:1 with Ammonia at nominal m/z 183 ion |
NCI scan sensitivity | 2-uL injection of a 100-fg/uL OFN standard scanning
from 50 to 300 u will give 600:1 S/N with Methane and 300:1 with Ammonia at nominal m/z 272 ion |
Physical Requirements
(with Agilent 6890 and 7890A GCs)
Dimensions (GC/MSD) | 88 cm (w) x 56 cm (d) x 50 cm (h)
Additional space should be added for the data system and printer. |
Weight (GC/MSD) | 81 or 96 kg (depending on configuration) |