Technical Specifications
Technical Specifications of produced water:
- Resistivity/Conductivity:
> 0.1 MOhm x cm to > 5 MOhm x cm
< 10 µS/cm to < 0.2 µS/cm
- production rate:
8 L/h to 40 l/h
- Total Organic Carbon (TOC):
< 20 ppb to 50 ppb
Technical Specifications |
||||
---|---|---|---|---|
Model | AquaRO² | AquaRO²DI | AquaRO² 8 P10 | P10 DI | EDI² |
pure water quality | < 10 µS/cm | < 2 µS/cm; Type II | < 10 µS/cm | < 2 µS/cm | < 0.2 µS/cm; Type II |
resistivity | > 0.1 MegaOhm/cm | > 0.5 MegaOhm/cm | > 0.1 MegaOhm/cm | > 0.5 Megaohm/cm | > 5 MegaOhm/cm |
total organic carbon (TOC) | < 50 ppb | < 40 ppb | < 40 ppb | < 20 ppb |
production rate | 8 L/h to 40 L/h | 8 L/h to 20 L/h | 8 L/h | 10 L/h to 40 L/h |
bacteria | < 50 cfu/mL | < 50 cfu/mL | < 50 cfu/mL | < 10 cfu/mL |
DNAse | < 5 pg/mL* | |||
dimensions, weight, power (except AquaRO² 40 and EDI²40) | ||||
AquaRO²: 504 x 340 x 535 mm, 16 to 20 kg, 110 V - 230 V | ||||
EDI²: 504 x 340 x 535 mm, >40 kg, 110 V - 230 V |
Feed Water Requirements |
||
---|---|---|
Model | AquaRO² | EDI² |
type of feed water | potable water / tap water | potable water / tap water |
feed water conductivity | < 1400 µS/cm | < 1400 µS/cm |
inlet pressure | 1.5 to 6 bar | < 5 bar |
free chlorine | < 0.1 mg/L | < 0.02 mg/L |
Silt Density Index (SDI) | < 3 | < 3 |
pH | 3 to 9 | 3 to 9 |
temperature | 5 to 25 °C | 5 to 25 °C |
iron / manganese | n. a. / n. a. | < 0.01 mg/L / < 0.01 mg/L |
sulfate | n. a. | < 2 mg/mL |