Weight | 1600 oz |
---|---|
Dimensions | 40 × 48 × 48 in |
Model | HCP50 (56L), HCP105 (107L), HCP150 (156L), HCP240 (241L) |
Stainless Steel
Scratch resistant, hygienic and durable.
Humidity chambers are used to simulate a wide variety of humidity conditions and their effects on the product or material in question. Simulating humidity without condensation is necessary for testing in all sorts of industries from zoology to electronics. Memmert humidity chambers provide high-precision temperature and humidity control that ensures reliable, reproducible, and documentable testing data without producing any condensation inside the chamber.
Memmert humidity chambers feature unique stainless steel that is scratch resistant, hygienic, robust, and ever-lasting. AtmoCONTROL allows for even the most complex ramp programs to be created in minutes with a mouse or touchpad on your notebook. Plus, a streamlined, easy-to-read front display allows for simplified monitoring of the chamber’s conditions during testing. We offer humidity chambers in all sorts of shapes, sizes, and dimensions so you’re sure to find one that fits your space and requirements. Please note this chamber comes standard with perforated shelves.
Scratch resistant, hygienic and durable.
Twice the features of SingleDISPLAY. Monitor programs easily via Ethernet or USB.
Create & monitor programs.
A limited 3 year warranty is included with the option to purchase an extra year.
Active humidity control from 20% to 95% rh offers unequaled uniformity of temperature and humidity within the chamber.
The ideal chamber for environmental testing, environmental simulation, accelerated service life tests, and 85/85 tests according to IEC 60068-2-67 and IEC 60068-2-78.
Memmert’s Humidity Chambers (HCP) provides the optimum homogeneity of humidity and temperature.
Technicians can independently set temperature and humidity values on the same ramp cycle for the exact environmental simulation.
Appliance size | 50 | 105 | 150 | 240 | |
Appliance width D* [mm] | 559 | 719 | 719 | 759 | |
Appliance height E* [mm] (varies due to adjustable feet) | 795 | 851 | 1071 | 1181 | |
Appliance depth F* (without door handle) [mm] | 521 | 591 | 591 | 691 | |
Depth of door handle [mm] | 56 | ||||
Chamber width A* [mm] | 400 | 560 | 560 | 600 | |
Chamber height B* [mm] | 425 | 480 | 700 | 810 | |
Interior depth C* [mm] (less 35 mm for fan) | 330 | 400 | 400 | 500 | |
Chamber volume [litres] | 56 | 107 | 156 | 241 | |
Net weight [kg] | 55 | 75 | 90 | 110 | |
Weight including packaging [kg] | 74 | 100 | 116 | 145 | |
Power [W]
|
115 V, 50/60 Hz | 1520 | 1720 | 1800 | 1840 |
230 V, 50/60 Hz | 1520 | 1720 | 1800 | 1840 | |
Current consumption [A]
|
115 V, 50/60 Hz | 13,2 | 15 | 15,7 | 16 |
230 V, 50/60 Hz | 6,6 | 7,5 | 7,8 | 8 | |
max. number of sliding grids/shelves | 5 | 6 | 10 | 12 | |
max. load per sliding grid [kg] | 15 | ||||
max. load per appliance [kg] | 75 | 90 | 120 | 140 | |
Operating temperature range | 7 °C above room temperature up to 90 °C | ||||
Setting temperature range [°C] | 18 to 90 | ||||
Adjustment precision [°C] | 0.1 | ||||
Adjustment range of active humidity control [% rh] (only for appliances with the corresponding configuration) |
20 to 95 | ||||
Setting accuracy humidity [%] | 0.5 | ||||
* See Figure |
The more complex twin display allows users to control every aspect of equipment functionality. Even complex processes such as multiple overtemperature protection, visual and acoustic alarm as well as automated alarm messages can be programmed quickly and easily.
Weight | 1600 oz |
---|---|
Dimensions | 40 × 48 × 48 in |
Model | HCP50 (56L), HCP105 (107L), HCP150 (156L), HCP240 (241L) |