Replacement Parts & Accessories
To view available parts for this instrument and instruments like it, click here.
Our parts and accessories inventory is constantly changing. Don't see what you're looking for? Give us a call, we'll be happy to help.
Relative humidity levels up to 90% are achieved in the growth chamber through the use of a 3-liter reservoir external to the chamber. The chamber air is routed through the water vapor saturated the air in an evaporate box to add humidity when the RH sensor indicates more humidity is needed to meet set point.
A single source dual wave infrared (IR) sensor controls CO2 levels within the growth chamber to maintain an optimal PH. The wavelengths used are absorbed only by CO2, making the measurements insensitive to other components such as water vapor. Advanced design provides a stable output minimizing drift, and requires less frequent calibrations.
The Direct heating elements or the water jacket are wrapped in high density insulation which surrounds the growth chamber. Dual temperature sensor probes relay information back to the NuTouch Electronic Control System and make adjustments as needed.
Est. Shipping Width Inches (mm) | 31 (787) |
Est. Shipping Height Inches (mm) | 51 (1219) |
Est. Shipping Depth Inches (mm) | 31 (787) |
Access Port Location | Right Side |
Air Cleanliness (ISO 14644) | ISO-Class-5 |
Chamber Air Circulation Method | Air Pump |
Chamber Air Filter | 99.97% Efficiency @ ≥ 0.3 Micron Capsule Filter Located in Sensor Bay |
Chamber Heater(s) Location | Walls, Top and Bottom of Chamber |
Chamber Jacket Technology | Air Jacket via One or Two Axial Fans Drawing Air into Shell |
CO2 Accuracy | ± 0.1% |
CO2 Control Logic | Fixed Algorithm/Manual, Environmental Adaptable |
CO2 Display Resolution | 0.1% |
CO2 Range | 0.1 to 20%, 5% Default Set Point |
CO2 Recovery | Up to 5% ± 0.2% / -0.5 in 5 minutes Average |
CO2 Sensor Type | Infrared (NDIR) |
Control System | NuTouch |
Data Output Features | RS-485, USB, 4-20mA |
Door and Perimeter Heater Control Logic | Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable |
Electrical Configuration | 115V50Hz, 230V5060Hz |
Exterior Depth Inches (mm) | 27.188 (691) |
Exterior Door Seal | Magnetic Outer Door Gasket |
Exterior Height Inches (mm) | 39.688 (1008) |
Exterior Material | Cold Rolled 16 & 18 Gauge Steel, Finished in Powder Coated Polyurethane Finish |
Exterior Width Inches (mm) | 26.750 (680) |
Footprint Depth inches (mm) | 20.75 (527) |
Footprint Width inches (mm) | 23.75 (603) |
In-line Gas Filters | 99.97% Efficiency @ ≥ 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor |
Segmented Doors | Option, Set of 6 Inner Doors, 2 Across by 3 Down |
Inner Doors | One |
Insulation | High Density Insulation with High “R” Rating |
Interior Depth Inches (mm) | 20.053 (509) |
Interior Door Seal | Silicone Rubber Gasket |
Interior Height Inches (mm) | 28.500 (724) |
Interior Material | Polished Type 304L 16 Gauge Stainless Steel |
Interior Volume Cubic Feet (Liters) | 7.06 (200) |
Interior Width Inches (mm) | 21.257 (540) |
Net Weight lbs (kg) | 235 (106) |
O2 Accuracy | ± 0.25% |
O2 Control Logic | Fixed algorithm/ manual environmental adaptable |
O2 Display Resolution | 0.1% |
O2 Measuring Range | 0 – 25% |
O2 Set Point Range | 0.5 -21.0% (ambient) (21.0% Default) |
O2 Recovery | 5.0 ± 0.5% twenty minutes on Average |
O2 Sensor Type | Zirconia Ceramic |
Power Cord Length Foot (Meter) | 8 (2.5) |
Remote Alarm Terminals | Standard |
RH Accuracy | +5% /- 3% |
RH Control Method | Water Pan or Evaporator Box, Depending on Model |
RH Control Logic | Fixed algorithm/manual environmental adaptable |
RH Control Range | Ambient dependent to 90% |
RH Display Resolution | 0.1% |
RH Evaporator Box Capacity | 3L |
RH Recovery | 90% +5%/-3% within 25 minutes on average |
RH Sensor Type | Precision capacitance bulk-polymer |
RH Water Pan Capacity | 2L (Recommended Fill 1.5L) |
Sample Port Location | Inner Glass Door |
Shelf Capacity lbs (kg) | 25 lbs per Shelf / 125 lbs per Incubator |
Shelf Depth Inches (mm) | 18 3/4 (476) |
Shelf Material | Stainless Steel or Copper Alloy |
Shelf Width Inches (mm) | 19 3/4 (502) |
Shelves | 4 Shelves Supplied, 20 Shelves Maximum Capacity |
Est. Shipping Weight lbs (kg) | 325 (147) |
Decontamination Cycle | 95°C Humidified Decontamination, 145°C Dry Decontamination Cycle |
Temperature Accuracy °C | ± 0.1% |
Temperature Control | Direct Heat |
Temperature Control Range | 5°C above ambient (to a 30°C max ambient) to 55°C |
Temperature Display Resolution | 0.1°C |
Temperature Recovery | 0.12°C/min. on Average |
Temperature Sensor Type | Precision Integrated Circuit |
Temperature Set-Point Range | 5°C to 55°C (37.0°C Default) |
Temperature Uniformity °C | ± 0.35°C @ 37°C |
Water Pan Depth | 1.5 (38) |
Water Pan Length Inches (mm) | 12 (305) |
Water Pan Width | 10 (254) |
NU-5841 | NU-5841E | |
---|---|---|
Electrical Configuration | 115V60Hz | 230V5060Hz |
To view available parts for this instrument and instruments like it, click here.
Our parts and accessories inventory is constantly changing. Don't see what you're looking for? Give us a call, we'll be happy to help.
Service Agreements |
Silver |
Gold |
First Class Ship & Fix |
Economy Ship & Fix |
Preventative Maintenance Visit | 1 Preventative Maintenance Visit per Year | 1 Preventative Maintenance Visit per Year | ||
Parts Covered | ||||
Labor Covered | ||||
Unlimited Remote Tech Support | ||||
Depot Repair | ||||
Depot Preventative Maintenance |
GMI was a very professional business to work with. Shipping was quick and easy. Technical Support was quick to respond. I would highly recommend GMI for all your laboratory equipment needs.